Introduction:HY gas pressure calibration equipment consists of hand pump, pressurized pump, shut-off valve, back check valve and pressure output interface to form an overall curing structure, the product structure is reasonable, the individual lightweigh……

HY gas pressure calibration equipment consists of hand pump, pressurized pump, shut-off valve, back check valve and pressure output interface to form an overall curing structure, the product structure is reasonable, the individual lightweight, with good sealing performance, boosting the pressure smoothly, boosting the pressure quickly, labor-saving, easy to operate and so on, it is the piston pressure source is ideal for the new generation of products.
1 Hand pump
2 Secondary pressurized pump
3 Diverter valve
4 Block valve (right valve)
5 Pressure output port
6 Check valve (left valve)
1、Pressure range: -0.095~4MPa -0.095-6MPa -0.095-10MPa
2、Stability:0.05%F-S
3、Working medium: air
4、Output interface: M20*1.5 inside
5、Exterior size: 430mm*300*mm*330mm
First Hand pump
Repeatedly lift and press to drive the piston of the pump body to realize boosting pressure
Second the second pressurized pump
1、Swing the regulator in to raise the pressure value, screw out to lower the pressure value;
2、After the pressure is stabilized, it can be used as a "fine-tuning valve";
Third the reversing valve
Push to the left for positive pressure, push to the right for negative pressure;
Fourth the stop valve
Positive pressure to 1.6 MPa or more, increase the pressure to 6 MPa when closing the shut-off valve, stable performance is ideal;
Fifth the check valve
1、Must be closed when getting pressure;
2、Swing open to quickly reduce the pressure value.
1、access to the standard meter and the instrument being calibrated should be free of water and corrosive substances and other impurities, otherwise it will affect the performance and service life of the equipment;
2、in unloading the pressure gauge, the system must be in zero pressure state;
3、the pressure source is strictly prohibited and other oil: mixed use;
4、pressurization, such as not go up, the reason is the quick connector "O" ring damage, should be replaced in a timely manner on the 9 * 2.65 "O" ring.
TOP